. Electromagnetic immunity

Guidance and Manufacturer's Declaration – Electromagnetic Immunity
ARTROMOT®-S is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the ARTROMOT®-S unit is
used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic
± 6 kV
discharge (ESD) to
contact
IEC 61 000-4-2
± 8 kV air
Electrical fast
± 2 kV for
transient/bursts
power supply lines
to
± 1 kV for input/
IEC 61 000-4-4
output lines
Surges to
± 1 kV
IEC 61 000-4-5
line-to-line voltage
± 2 kV
line-to-earth
voltage
Voltage dips, short
< 5 % U
interruptions and
(> 95 % dip in U )
voltage variations
for 1/2 cycle
on power supply
40 % U
input lines to
(60 % dip in U )
IEC 61 000-4-11
for 5 cycles
70 % U
(30 % dip in U )
for 25 cycles
< 5 % U
(> 95 % dip in U )
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field to
IEC 61 000-4-8
NOTE: U is the a.c. mains voltage prior to application of the test level.
Compliance level
Electromagnetic
environment - guidance
± 6 kV
Floors should be wood, con-
contact
crete or ceramic tile. If floors
± 8 kV air
are covered with synthetic
material, the relative humidity
should be at least 30%.
± 2 kV for
Mains power should be that of
power supply lines
a typical commercial or hospi-
± 1 kV for input/
tal environment.
output lines
± 1 kV
Mains power should be that of
line-to-line voltage
a typical commercial or hospi-
± 2 kV
tal environment.
line-to-earth
voltage
< 5 % U
Mains power should be that of
(> 95 % dip in U )
a typical commercial or
for 1/2 cycle
hospital environment.
40 % U
If the user of the
(60 % dip in U )
ARTROMOT®-S unit requires
for 5 cycles
continued operation during
70 % U
power mains interruptions, it is
(30 % dip in U )
recommended that the
for 25 cycles
ARTROMOT®-S unit be
< 5 % U
powered from an uninterrupti-
(> 95 % dip in U )
ble power supply or a battery.
for 5 s
3 A/m
Power frequency magnetic
fields should be at levels cha-
racteristics of a typical location
in a typical commercial or
hospital environment.