DJO Global ARTROMOT-S3 Série Mode D'emploi page 68

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 90
– Timer:
– Speed:
– Reverse on load -
motor A:
– Reverse on load -
motor B:
– Motor A:
– Motor B:
– Synchronized mode:
– Total therapy time:
– Special functions:
LEVEL :
n Reverse on load - motor A
(safety circuit)
The motors automatically switch to op-
posite directions of movement when the
patient's resistance (load) exceeds the
set value.
Adjustable levels for reverse on load fea-
ture: 1 - 25. At level 1, very low resistance
will cause the device to reverse; at level
25, a high resistance is required to initi-
ate the reversal.
Default: level 25
n Reverse on load - motor B
(safety circuit)
The motors automatically switch to op-
posite directions of movement when the
patient's resistance (load) exceeds the
set value.
Adjustable levels for reverse on load fea-
ture: 1 - 25. At level 1, very low resistance
will cause the device to reverse; at level
25, a high resistance is required to initi-
ate the reversal.
Default: level 25
Caution!
Patient hazard —
The reverse on load feature is a safety
measure to protect the patient in the
event of cramps, spasms, locked joints
and similar situations. The manufac-
turer cannot be held liable for misuse of
this feature.
continuous
n Motor A ON/OFF
operation
To permit fully isolated movements,
100 %
the motors can individually be turned
on and o . Motor A e ects adduction
and abduction, motor B e ects internal
25
rotation and external rotation.
For an isolated internal / external
25
rotation movement, set motor A to the
ON
programmed position (abduction/ad-
duction), then turn the motor o .
ON
In normal mode the display indicates the
ON
symbol "OFF" during therapy, instead of
0
the programmed adduction/abduction
angles.
disabled
Default: motor A ON
n Motor B ON/OFF
To permit fully isolated movements,
the motors can individually be turned
on and o . Motor A e ects adduction
and abduction, motor B e ects internal
rotation and external rotation.
For an isolated adduction/abduction
movement, set motor B to the pro-
grammed position (internal / external
rotation), then turn the motor o .
In normal mode the display indicates the
symbol "OFF" during therapy, instead of
the programmed internal/external rota-
tion angles.
Default: motor B ON
Note!
Please note that one motor (A or B)
must always be turned on. Otherwise
the following symbol will be displayed
when you press START :
LEVEL :
n Transport setting
With this function, the carriage will
move to a position optimally suited
for packing the CPM device. Select the
function and press the START key. The
carriage moves to the transport posi-
tion. In the top left-hand corner of the
display you will see
section 6.3 Transport).
(also refer to

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Artromot-s3 standard80.00.023Artromot-s3 comfort80.00.024

Table des Matières