DJO Global ARTROMOT-S3 Série Mode D'emploi page 76

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 90
. . Isolated elevation
(flexion)
1. First adjust an angle of 90° for external
rotation and deactivate motor B as
follows:
a. Press the MENU key on the program-
ming unit to access level 1 (M1).
b. Press the parameter key for external
rotation
and move the car-
riage to the rotation position of 90°,
using the + / - keys.
Note!
– For an exclusive elevation protocol,
motor B for the rotation movement
must be turned o as described in
steps c to e below.
– Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
c. Press the MENU key on the program-
ming unit to access level 3 (M3).
d. Press the Motor B ON/OFF
parameter key to activate the
parameter.
e. Deactivate motor B by pressing the
parameter key Motor B ON/OFF again
or the „–" (minus) key. The check mark
in the circle next to the symbol must
have disappeared.
f. Save the entered values by press the
STOP key.
2. Complete the mechanical settings and
adjust the carriage to the patient as
follows:
a. Adjust an angle of 90° for horizontal
adduction/abduction
ing knob (12).
b. Secure and hold the motion element
at the holding tube for length adjust-
ment and open the locking screw
(14) for height adjustment
a height at which the pivot points of
motor A and the shoulder are at the
same level.
c. Now open clamping lever (18) for ad-
justment of the upper arm length
locking screw (20) for adjustment of
è
the elbow angle
(28) for adjustment of the forearm
, using index-
Œ
. Select
Å
,
and locking screw
rest inclination and set the motion
element to a position that is comfort-
able to the patient and suitable for
the therapy session . When done,
tighten all screws and levers.
d. Open clamping lever (26), adjust the
^
forearm length
close the clamping lever.
3. In the next step, program the elevation
range of motion as follows:
a. Press the MENU key on the program-
ming unit to access level 1 (M1).
b. Next set the desired range of mo-
tion for adduction and abduction by
first pressing the parameter key for
adduction
with the +/- keys.
c. Set the abduction
same way.
Note!
Having programmed the ranges of
motion, you can program additional op-
tions, such as pause or speed.
4. Save the settings with the STOP key,
set the carriage to the starting position
with the START key and initiate therapy
by pressing the START key again.
to the patient and
and set the values
value in the