Dentsply Sirona SmartLite Focus Mode D'emploi page 15

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 20
Disinfection:
Manual (Inter-
mediate-Level)
Packaging
Sterilization
Drying
Maintenance
Storage
Manufacturer
Contact
Incidental surface contact of the power supply and cord with water, soap or a water-
based hospital-level disinfection solution will not damage the material of construction.
Do not allow any solution to penetrate the casing.
1. After cleaning wipe all device surfaces with a single use cloth
in combination with an alcohol-based, tuberculocidal, quaternary
ammonium solution (e.g., BBraun Meliseptol
2. Thoroughly disinfect the surface with the soaked cloth, paying
special attention to any cavities and holes. Ensure the entire
surface remains wet for the entirety of the required contact time
in accordance with the disinfectant solution manufacturer's In-
structions for Use.
3. Rinse for 30 seconds under running potable water.
4. Dry with a lint-free single use cloth.
Paper/plastic steam sterilization pouches (e.g., AssurePlus
lization Pouches) may be used, but are not required.
After Cleaning and Disinfecting, steam autoclaving is required.
Prevacuum Steam Sterilization:
• Full Cycle: 134 °C/274 °F for 3 minutes 30 seconds.
Follow manufacturer's instructions for loading and operation cy-
cle.
Use the drying cycle of the autoclave, minimum 5 minutes.
Allow the components to fully air dry before storage.
Visually inspect to ensure that all contamination has been re-
moved.
If the shield is discolored, damaged, worn, or distorted it should
be discarded.
No additional maintenance or lubrication is recommended.
Store the sterilized light shield at room temperature, away from
moisture or excessive humidity.
Instruments steam autoclaved bagged should remain bagged un-
til ready for use. Prior to subsequent re-use, inspect the steriliza-
tion pouch and the shield. If the integrity of the sterilization pouch
has been compromised, the shield must be reprocessed prior to
use. Instruments steam autoclaved unwrapped should be used
immediately.
Light handpiece and charging base should be cleaned, disinfected
and dried as outlined in the section above prior to storage.
Within the United States, call Dentsply Sirona at 1-302-422-4511.
For areas outside the United States, contact your local Dentsply
Sirona representative.
Rapid).
®
Steri-
®
11

Publicité

Table des Matières
loading

Table des Matières