Instructions for Cleaning and Disinfecting
SmartLite
Focus
®
®
Warnings
Limitations on
reprocessing
Point of use
Containment  /
Transportation
Light Handpiece, Charging Base
• The SmartLite
Focus
®
disinfected/sterilized as outlined below. The SmartLite
light handpiece and charging base are not sterilizable by auto-
clave.
• Intermediate-level disinfection is appropriate for the handpiece
and charging base.
• Do not autoclave in steam autoclave.
• Do not clean/disinfect in automated washer/disinfector.
• Do not immerse in liquid.
• Disconnect the power supply plug from the power outlet and
charging base unit prior to cleaning/disinfection.
• Repeated reprocessing has minimum effect on these instru-
ments. End of life is normally determined by wear and damage
due to use.
• Device can be reprocessed for up to 6000 times.
• Cold liquid immersion disinfection/sterilization, chemical vapor
sterilization, and dry heat sterilization methods have not been
tested or validated for efficacy and are not recommended for
use.
• Remove protective barrier and discard according to local regu-
lations.
• Remove excess soil with disposable cloth/paper wipe in com-
bination with an alcohol-based, tuberculocidal, quaternary
ammonium solution with a label claim for cleaning e.g., Volo-
Wipes
disinfecting/cleaning/deodorizing Wipes or pH-neutral,
®
phosphate-free cleaning solution (e.g., Dr. Schumacher Instru
Plus [3%]).
• Remove SmartLite
below.
• Do not remove battery pack. Do not attempt disassembly of light
handpiece or charging base.
• It is recommended that the device be reprocessed as soon as is
reasonably practical following use.
• Start reprocessing within 1 hour after use.
• Charging base should be reprocessed as soon as reasonably
practical after being exposed to spatter or spray of body fluids,
or touched by contaminated hands or contaminated light hand-
piece.
No particular requirements. Handle with care.
Shield should be removed and cleaned  /
®
Focus
Light Shield. Reprocess as outlined
®
®
Focus
®
®