To Use The Warm Up Key - Kinetec Breva V01.0 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

11.12.

To use the WARM UP key

Beginning
To stop the unit
To select the
Plantar Flexion/Dorsiflexion movement
To select Warm Up mode
OR
To select the
Inversion/Eversion movement
To select the right or left foot
To select Warm Up mode
To start the movement
Warm Up rules.
The device Kinetec Breva™ starts at 70% of the full ROM, increasing 5% of the range each full
cycle until the pre-set ROM is reached.
Note:
- The parameters should only be modified while the unit is stopped.
- Pauses and ByPass mode are not available during warm up cycles.
- The Warm Up cycles are only performed when the movement is first started.
- To start a new Warm Up session, press the "Warm Up" key twice.
- The calculation mode used enables the pre-set ROM to be reached in approximately seven full
cycles.
Example:
For a Warm Up treatment with a pre-set ROM from 40° to 30°.
The first cycle starts at 32° plantar to 22° dorsal, increasing 5% of the range each cycle.
Kinetec Breva™
Keys to
press
or
30°
30°
-5°
-5°
-40°
-40°
Display
PL/FLEX DOR/FLEX
20 STOP 10 5
PL/FLEX DOR/FLEX
20 STOP 10 5
PL/DOR FLX W.UP
20 STOP 10 5
INVERSION/EVERS.
RIGHT:+
LEFT:-
EV Right INV
10 STOP 10 10
EV W UP RIGHT INV
10 STOP 10 10
EV W UP RIGHT INV
10 RUN 10 10
EN
Remarks
Check that the hand control is not
locked (see page 7).
To set movement values, see page 8.
The angle display changes with the
current movement.
12/20