Checking of the Equipment Before Each Use
Each time the use of the electrosurgical equipment is planned a check of the most impor-
tant safety aspects has to be implemented considering at least the following:
Check the integrity of cords, connections, wires breakage, etc.
Assure that all the electrical equipment is properly grounded.
Assure that all the accessories that should be used are available and sterilized.
Check, by activating the power switch, the functioning of the emission sounds
warnings.
Function and Safety Check and Test
At least once a year, the biomedical engineering department or other qualified personnel
should do the following check and test:
Check of the connectors and mains supply cord conditions.
Visual check of the mechanical protections.
Check of the protections against the danger due to liquid's pouring, dripping,
EN
moisture, liquid's penetration, cleanliness, sterilization and disinfection.
Check of the Equipment's Data on the Label.
Check of the availability of the Instruction's Manual.
Check the functioning of the H.F. output controls.
Check the uniformity of the resistance through the surface of the patient plate.
Test the earth conductivity resistance.
Test the earth leakage current.
Test H.F. leakage current.
Control of the neuromuscular stimulation.
Control of the accuracy of the output power.
EN 20