Adjustment Of The Acoustic Signal Level; Connectors; Back Panel - Carlo De Giorgi SURGERY PLUS Mode D'emploi

Bistouri electrique pour chirurgie mono polaire
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Adjustment of the Acoustic Signal Level
To modify the emission acoustic signal it is necessary to follow those indications:
1.
Switch on the unit through the mains switch while the CUT pushbutton is
maintained pressed.
2.
When the unit has finished to check internal parameters, on the display ap-
pears the message S + value of the preset level (example S3). Now, the CUT push-
button can be released.
3.
Through the knob it is possible varying the emission acoustic level. During
the variation the sound emitted by the unit corresponds to the preset level.
4.
Press the CUT pushbutton to confirm the level.
EN

Connectors

Neutral Electrode Connector
Handle Connector
Foot-switch Connector
Back Panel
EN 16
Level
Sound emission until 1m dis-
tance from the frontal panel
S1
55 dBA
S2
60 dBA
S3
65 dBA
S4
70 dBA
S5
75 dBA
This is the point of connection of the neutral elec-
trode.
This is the point of connection of electrode handle.
On the left part of the front panel it is foot-switch con-
nector.
1 Fuses holder/Voltage selector
2 Power On-Off switch
3 Mains voltage connector
4 Equipotential connector
MA505a_IT-EN-ES-FR-DE

Publicité

Table des Matières
loading

Table des Matières