Repairs - Carlo De Giorgi SURGERY PLUS Mode D'emploi

Bistouri electrique pour chirurgie mono polaire
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Problems
The unit doesn't respond to
the command of activation.
Error Code 91
Error Codee 92
Error Code 93
Error Code 94
Error Code95
Error Code 99

Repairs

High frequency cables and electrode holder handle cannot be repaired. Always substi-
tute a damaged part with a new one.
Fuses Substitution
Before substituting the fuse, disconnect the unit from the mains system
Only use fuse of the kind 5x20, proceed as follows:
(A-B) Extract the fuse holder drawer from the power module.
(C) Insert the fuses making reference to the following chart:
SURGERY PLUS
(D) Extract and rotate the detachable part in way to read the
correct voltage in the (E) window – reinsert the fuse holder in the
module.
MA505a_IT-EN-ES-FR-DE
Probable Cause
Breakdown of the pedal.
Wrong connection of the ped-
al.
Current delivery control acti-
vated during switching on.
Error in the management
board.
Error in the management
board.
Error in the data conversion
circuit.
Error of the reference voltage
value.
Error in the output power
activation circuit.
Mains Voltage
110-120 V Delayed Fuse 2x T2 A / 5 x 20 mm
Mains Voltage
220-240 V Delayed Fuse 2x T1 A / 5 x 20 mm
Solutione
Replace the pedal.
Disconnect the pedal and
switch on the unit again.
Call for Service.
Call for Service.
Contattare il Servizio di As-
sistenza Tecnica.
Verify the main voltage.
Call for Service.
Call for Service.
EN 19
EN

Publicité

Table des Matières
loading

Table des Matières