Télécharger Imprimer la page

B.Braun Aesculap DX646T Mode D'emploi page 4

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 10
Manual cleaning/disinfection
Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.
After manual cleaning/disinfection, check visible surfaces visually for residues.
Repeat the cleaning /disinfection process if necessary.
Manual cleaning with immersion disinfection
Phase
Step
T
[°C/°F]
I
Disinfecting clean-
RT
ing
(cold)
II
Intermediate rinse
RT
(cold)
III
Disinfection
RT
(cold)
IV
Final rinse
RT
(cold)
V
Drying
RT
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Drain any remaining water fully.
Phase III
Fully immerse the product in the disinfectant solution.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible surfaces).
Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.
Rinse lumens with an appropriate disposable syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
T
[°C/°F]
I
Ultrasonic cleaning
RT
(cold)
II
Intermediate rinse
RT
(cold)
III
Disinfection
RT
(cold)
IV
Final rinse
RT
(cold)
V
Drying
RT
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible
surfaces are immersed and acoustic shadows are avoided.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Drain any remaining water fully.
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
1
-
D–W
-
15
2
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
1
-
FD-W
-
-
-
-
-
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
D–W
-
15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
FD-W
-
-
-
-
-
Phase III
Fully immerse the product in the disinfectant solution.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.
Phase IV
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.
Rinse lumens with an appropriate disposable syringe at least five times.
Drain any remaining water fully.
Phase V
Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.
Mechanical cleaning/disinfecting
Note
The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
T
[°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
III
Intermediate rinse
>10/50
IV
Thermal disinfecting
90/194
V
Drying
-
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)
*Recommended: BBraun Helimatic Cleaner alkaline
Check visible surfaces for residues after mechanical cleaning/disinfecting.
Mechanical cleaning/disinfection with manual pre-cleaning
Note
The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual pre-cleaning with a brush
Phase
Step
T
t
[°C/°F]
[min]
I
Disinfec-
RT
>15
tant
(cold)
cleaning
II
Rinsing
RT
1
(cold)
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.
Phase I
Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.
Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.
If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.
Mobilize non-rigid components, such as set screws, links, etc. during cleaning.
Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.
Phase II
Rinse/flush the product thoroughly (all accessible surfaces) under running water.
Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Manual pre-cleaning with ultrasound and brush
Phase
Step
T
[°C/°F]
I
Ultrasonic cleaning
RT
(cold)
II
Rinsing
RT
(cold)
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Stabimed
t
Water
Chemical/Note
quality
[min]
3
D–W
-
10
FD-W
Concentrate, alkaline:
– pH = 13
– <5 % anionic surfactant
0.5 % working solution
– pH = 11*
1
FD-W
-
5
FD-W
-
-
-
According to the program for cleaning
and disinfection device
Conc.
Water
Chemical
[%]
quality
2
D–W
Aldehyde-free, phenol-free, and
QUAT-free concentrate, pH ~ 9*
-
D–W
-
t
Conc.
Water
Chemical
[%]
quality
[min]
>15
2
D–W
Aldehyde-free, phenol-free,
and QUAT-free concentrate,
pH ~ 9*
1
-
D–W
-

Publicité

loading