Mettre en service avec la HMI intégrée
7.3 Aperçu des menus
Menu item
VOLUME
VOLUME
FLOW
CUSTOM
LOCL DISPLAY START VIEW
Menu item
SIGNAL
32
Parameter
SATURATE CUR
= Main Current
Output_Configuration:
Current Limits
FAULT CUR =
Main Current
Output_Configuration1:
Current Limits
FAIL SAFE LOE = Echo
Loss Detection: Mode
LOE TIMER
VESSEL SHAPE
VESSEL DIM A
VOL UNITS
UPPER SCALNG
PRIM MEASDEV
VFLOW UNITS = Level
Scaling: ScalingUnit
SCALE FORMAT
UPPER SCALNG
UPPER SCALNG
CUSTOM CURVE =
Linearization Points:
Level Value, Volume
Value
Parameter
SIG QUALITY
ECHO CONFIG
Setting
3,8 ... 20,5 mA, 4 ... 20 mA
≤ 3,6 mA, ≥ 21 mA
HOLD = off
FAULT = Failure
Echo Loss Detection: Echo Loss Time
LINR
Linear
CONIC
Conical Bottom
FLAT
Inclined bottom
CYLIN
Horizontal cylindrical vessel
USER
User programmable
Intermediate Height
Level Scaling: Scaling Unit
Level Scaling: Max Scaled Value
PBFLM
Flow Palmer-Bowlus flume
RWRC
Venturi-, trapezoidal weir,
rectangle weir
TPVWR
Flow VNotch-, triangle weir
CUSTM
User programmable
3
l/s, l/min, l/h, MI/d, m
/s, m
d, lb/s, lb/min, lb/h, gal/s, gal/min, gal/h,
3
3
gal/d, Mgal/d, ft
/s, ft
/min, ft
s, bbl/min, bbl/h, bbl/d, ImpGal/s, ImpGal/
min, Imp/Gal/h, ImpGal/d
Level Scaling: Decimal Places Display
Level Scaling: Max Scaled Value
Level Scaling: Max Scaled Value
X VALUE 1, Y VALUE 1, X VALUE 2, Y VALUE
2, ... X VALUE 32, Y VALUE 32
Display Config: Picture Selection
Setting
CONFIDENCE Level Reliability Value: Unit
ECHO SIG
Level Echo Info: Amplitude
STR
NOISE
Noise Detection Info: Noise
AVG
Level Abs
NEAR
Measuring Range
RANGE
ON
Active
Default
3,8 ... 20,5 mA
≤ 3,6 mA
HOLD
15 s
LINR
0 mm
l
100 l
PBFLM
3
3
3
/min, m
/h, m
/
l/s
3
3
/h, ft
/d, bbl/
0
100 l/s
100 l
0 %,
0 ... 100 %,
100 %
LEVEL
Default
CONFIDENCE
OFF
SITRANS LR150
Instructions de service, 08/2020,
A5E50352155