Télécharger Imprimer la page

Mechanical Functioning; Folding Arm; Tubehead; Extension Arm - GENDEX Oralix AC Manuel De Service

Masquer les pouces Voir aussi pour Oralix AC:

Publicité

Les langues disponibles

Les langues disponibles

ORALIX AC SYSTEM

4. MECHANICAL FUNCTIONING

4.1. FOLDING ARM

The spring tension of the Folding Arm is factory adjusted: as a result the arm with Tubehead shall return to
standby position, both horizontally and vertically, when slightly displaced by hand.
Otherwise the arm remains steady in work position when a larger displacement is applied.
If not, proceed first to an adjustment of the friction screws: when this is not sufficient adjust also the spring
tension.
Refer to section F for the relevant adjustment procedures.

4.2. TUBEHEAD

The Tubehead shall balance in all positions and the torque required to rotate it shall not exceed 1.8 Nm in the
horizontal plane, or 1.5 Nm in the vertical plane.
If an adjustment of the frictioning of the vertical rotation is required, refer to section F for the adjustment
procedure.

4.3. EXTENSION ARM

The Extension Arm is provided with friction screws for the adjustment of the horizontal rotation of the Folding
Arm.
Refer to section F for the adjustment procedure.

4.4. MOBILE STAND ADJUSTMENTS

Adjust if necessary the verticality of the Mobile Stand column as described in the relevant unit manual.
There is no facility on the Mobile Stand for the frictioning of the horizontal rotation of the
Folding Arm around the Mobile Stand column.
Gendex Dental Systems
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
NOTE
SETTING TO WORK
C-3

Publicité

loading