ORALIX AC SYSTEM
.
Doubling the surface of the steel plates respect to the size of the Wall Support will halve
the static load applied on the wall.
In case of other very weak walls (wood,etc.) a steel construction is needed.
The Timer is equipped with 4 expansion screws, which are suitable for all types of walls having a min. thickness
of 5 cm.
Drill the mounting holes (diameter 8.5 mm) in the desired location of the wall using the template BZ-2.
Gendex Dental Systems
Wall mounting by means of the "through the wall"
method. In case of other strong wall types (concrete
with R not Known, solid brick) it is recommended to
use the "THROUGH THE WALL" method of
attachment which makes use of a steel counter plate
(2 mm thick), of the same size of the Wall Support,
tied with screws or stay bolts (class ISO 8-8, diameter
8 mm), ensuring a tightening torque on the bolts of
10 N/m.
Wall mounting by means of 2 reinforcement steel
plates. In case of other less strong wall types (hol-
low tile, wood), which however can bear the specified
load without evident deformation, it is recommended
to use the "THROUGH THE WALL" method of
attachment, which makes use of an anterior rein-
forcement steel plate (2 mm thick) and a steel coun-
ter plate (2 mm thick) of dimensions larger than the
Wall Support, tied with screws or stay bolts (class
ISO 8-8, diameter 8 mm), ensuring a tightening torque
on the bolts of 10 N/m.
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
NOTE
INSTALLATION
B-3