Collimator; Timer; Weights - GENDEX Oralix AC Manuel De Service

Masquer les pouces Voir aussi pour Oralix AC:

Publicité

Les langues disponibles

Les langues disponibles

ORALIX AC SYSTEM
4.3. CO
Collimator type
Distance from Focal Spot to distal cone end
Irradiated field at distal cone end

4.4. TIMER

Versions
Selectable exposure times
Actual exposure time
"X-ray ON" indication
Preheating time
Compensation of mains fluctuations

4.5. WEIGHTS

Wall System
Mobile System
Tubehead Oralix AC, 230V
Folding Arm
Extension Arm
Wall Support
Densomat Timer
Secondent Timer
For the dimensions of the system see AZ-3 and AZ-4
·
·
among 19 settings, according to R10 scale:
0.03, 0.05, 0.06, 0.08, 0.10, 0.12, 0.16, 0.20, 0.26, 0.32,
0.40, 0.50, 0.64, 0.80, 1.00, 1.26, 1.60, 2.00, 2.50.
·
·
·
visual (LED) and audible (buzzer)
120 msec (default) 6 pulses @ 50 Hz, 7 pulses @ 60 Hz
by automatic correction of exposure time
45 cm
90 cm
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
metal cone, lead shielded, bayonet fitting
20 cm
6 cm diameter (circular, standard)
3 x 5 cm
(rectangular, optional)
Densomat, object programmed with +/- manual time
correction, with possibility to control 2 Tubeheads by
optional extension hardware.
Secondent, with manual time setting by time selec-
tor.
For 50 Hz is the number of X-ray pulses correspond-
ing to the selected time.
For 60 Hz is the number of X-ray pulses with mini-
mum deviation respect to the selected exposure
time.
the exposure time is measured between 25% of the
level of the anode current waveform.
The accuracy is +/-(10% +1 ms)
25 kg
60 kg
6.3 kg, without collimator
7.5 kg
3.5 kg
5.0 kg
1.5 kg
2.0 kg
2.0 kg
TECHNICAL DATA