INSTALLATION

3.3. ASSEMBLING THE MOBILE STAND

1
Place the two legs as shown by the picture.
WARNING
In order to simplify the installation, please use
the three wood-pieces from the wrapping as
shown in the drawing.
To insure that the two legs are mounted only in
the correct position, the two guide-bolts are
shaped in different thick.
Set the column on the pedestal, insert the two
2
exagonal screws M8 x 70 (a) and lock washers
supplied.
WARNING
The proper fixation of the column to pedestal is
fondamental for the strenght and stability of the
mobile stand.
By means of a proper exagonal wrench (b), tighten
3
firmly the column and the pedestal togheter.
Ensure that the timer mounting plate (c) of the
column is positioned at the rear of the Mobile
Stand, on the short legs side.