TECHNICAL DATA

4. TECHNICAL DATA

4.1. MAINS SUPPLY

Mains voltage
Current absorption
Mains fuse
Frequency
Line resistance

4.2. TUBEHEAD

Peak tube potential (see section D-3.2.)
Tube current (see section D-3.3.)
Radiation output
Tube rating chart (only 1 fixed working point)
X-ray tube
X-ray Tubehead
Minimum filtration permanently in the beam
Beam quality half value layer (H.V.L.)
Focal spot
Duty cycle
Minimum wait time
Radiation leakage
Leakage Tecnique factors
Anode angle
A-6
230 VAC +/-10%, single phase (polarized) or double phase
it is reccomended to use a dedicated magneto/thermal circuit breaker (10
A max) and an earth leakage circuit breaker (leakage current of 30 mA)
F6.3A / 250V, fast blow 2
50 Hz / 60 Hz ± 2 Hz
0.8 Ohm
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
nd
fuse for double phase, with by-pass jumper
60.0 kVp +/- 8%
(Nominal mains Voltage -10%)
65.0 kVp +/- 8%
(Nominal mains Voltage)
68.5 kVp +/- 8%
(Nominal mains Voltage+10%)
6. mA +/- 10%
Nominal mains Voltage -10%)
mA +/- 10%
Nominal mains Voltage)
. mA +/- 10%
Nominal mains Voltage +10%)
5 mGy/sec +/-30% at 20 cm from the focus
65 kVp, . 5 mA, 2.5 s
800-000 G , grid controlled, single focus
Oil immersed, shielded
2.0 mm Al equiv. @ 65 kVp (IEC 522/1976)
better than 1.7 mm Al @ 65 kVp
0.7 mm (IEC 336/1982)
1
3 sec
less than 4 microGy/h (5 mR/h) @ 1 meter
0.1 mA @ 65 kVp
19.5°
respect to the
+/- 0.5
ORALIX AC SYSTEM
p
axis
Gendex Dental Systems