7 Technical Data

Mains power supply:
Maximum mains voltage deviation:
Maximum frequency deviation:
Maximum current absorption:
Stand-by current absorption:
Mains resistance:
X-ray tube:
Secondary voltage, nominal:
Secondary current, nominal:
Maximum duty cycle ratio:
Permissible mean current:
Focus (focal spot):
Target material
lnherent filtration, minimum:
Output dose rate (approximate, indicative):
HVL - Half Value Layer (approximate, indicative):
SSD - Source (focus) to skin distance:
Manufacturer
Type:
X-ray field at tip of cone:
Leakage radiation:
(taking into account the 1:30 maximum duty factor)
Reach:
The serial number of the Gendex X-ray tube (PN 800-0003G3) is
stamped on the copper anode of the tube assembly.
Printed on: 22 Mär 2012, 09:22:12 am; Printed by: TOBIAS BAUER
GENDEX
230 V - 50/60 Hz.
± 10%
± 2 Hz
5 A
0.1 A
0.8
65 kVp
8.25 mA mean
1:30
360 mAs/hour
0.7 (IEC 336 - 1982)
Tungsten
2 mm Al equivalent (IEC 522 1976)
0.2 mGy/s @ 20 cm from source (at cone tip)
2 mm Al
20 cm
Gendex
800-0003G3
60 mm (round BLD)
35 x 45 mm (rectangular BLD)
less than 0.044 mGy/h @ 1 m from
source
1410 cm / 1860 cm
20
Oralix AC