GENDEX Oralix AC Manuel De L'utilisateur page 19

Masquer les pouces Voir aussi pour Oralix AC:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Default values (medium patient size, film speed coefficient =1):
Tooth mandible
: lower incisor
: lower canine
: lower premolar
lower molar
Bitewing
The lamp associated to a tooth, whose set time has been changed, lights on. Press key P to
save and enter Digital Mode Customization.
Digital Mode Customization
The message diG is displayed for 2 s, after that the display shows the digital mode coefficient (0.4s by
default). Lamp
is off and lamp P is flashing.
It is possible to set the timer for digital mode per default at start-up by following the instructions below.
Press the
key, the control LED lights up; press the P-key until "End" shows up in the display. As of
now the digital mode is set per default when starting up the timer. This setting can be made for one or
both tubeheads ( 1 and
original setting, press the
Using the s+ and s- keys, you may change the coefficient of the time scale as follows (according to the
R10 scale). The entire time scale is altered by shifting all nominal time values by the given number of
steps. By default, the exposure time scale is shifted by -4 steps for the Digital Mode (0.4s on display).
Press key P to save the selected coefficient, and to exit the Digital Mode Customization. The message
End is displayed. Switch off the timer, wait few seconds and switch on again to use the timer with the
customization.
This procedure must be completed as described above to became effective, and is concluded
only when "End" appears in the display.
Exposure Time
0.32 s
0.40 s
0.40 s
0.50 s
0.50 s
2) controlled by the timer, should these be connected. To restore the
key, the control LED does not light up.
Coefficient
0.32
Nr of steps
-5
Printed on: 22 Mär 2012, 09:22:12 am; Printed by: TOBIAS BAUER
GENDEX
Tooth maxilla
upper incisor
upper canine
upper premolar
upper molar
0.4
0.5
0.64
-4
-3
-2
15
Exposure Time
0.40 s
0.50 s
0.64 s
0.80 s
0.8
1
-1
-
Oralix AC

Publicité

Table des Matières
loading

Table des Matières