CORRECTIVE MAINTENANCE

3.2. FOLDING ARM SPRING TENSION

To adjust spring tension of Folding Arm, proceed as follows:
Position tubeside section of Folding Arm in high-
1
est position, parallel to floor; other section of Fold-
ing Arm should be vertical.
2
Remove snap out caps (a) from arm covers.
3
Push out the pins (b). Use special drift.
Partly raise covers (c) to access to the threaded
4
bushing.
Turn bushing (d) clockwise to increase spring
5
tension, counter-clockwise to decrease tension.
Use special drift.
6
Replace covers, pins and caps.

3.3. TUBEHEAD FRICTION

To adjust Tubehead friction at pivot axis, proceed as follows:
Remove snap-out disk (e).
1
Remove three adjustment screws (f) and check
2
for wear.
Reinstall screws (f) and adjust up to desired
3
Tubehead frictioning.
Apply loctite to screw heads.
4
Verify Tubehead proper balance and tension.
5
The specified Tubehead rotation torque
is 1.6 to 2.2 Nm.
6
Replace snap out cover disk.
Gendex Dental Systems
NOTE