ACteon SOPROLIFE Manuel D'utilisation page 11

Masquer les pouces Voir aussi pour SOPROLIFE:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 31
Guide and declaration of the manufacturer - electromagnetic immunity
SOPROLIFE device is intended to be used in the electromagnetic environment specified below.
CEI 60601
Immunity trial
Severity level
Conducted RF
3 Vrms
EN 61000-4-6
150 kHz
to 80 MHz
Radiated RF
3 V/m
EN 61000-4-3
80 MHz to 2.5
GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency band applies.
Note 2: These recommendations may not apply in every situation. Electromagnetic wave propagation is
modified by the absorption and reflection due to the structures, objects and persons.
E N G L I S H
The user should make sure it is used in this environment.
Compliance
level
3V
3V/m
Electromagnetic environment
Guide
Portable and mobile RF communication devices
should not be used at a distance from SOPROLIFE
device including the cables, lower than the
recommended separation distance, calculated
with the applicable formulas depending on the
emitter frequency.
Recommended separation distance
d = 1,16 √P
d = 1,16 √P 80 MHz to 800 MHz
d = 2,33 √P 800 MHz to 2.5 GHz
where
is the maximum rated output of the
transmitter in watts (W) by the transmitter
manufacturer and d the recommended separation
distance in metres (m).
The field levels emitted by the fixed RF
transmitters, determined by an electromagnetic
measurement of the site
the compliance level in each frequency band.
Interference may occur in the vicinity of the
devices bearing the following symbol:
S O P R O L I F E • U s e r ' s m a n u a l
a
, should be lower than
9

Publicité

Table des Matières
loading

Table des Matières