4. Put the sample cell in the center opening of the second row of the
heating block.
5. Select a temperature program of 150 °C for 60 minutes using the
keys, then push OK.
6. When the heating block is at the selected temperature (two beeps
are heard), the thermometer temperature should be the same as the
temperature shown on the display.
Troubleshooting
Refer to
Table 4
for error messages, possible causes and corrective
actions.
Table 4 Error messages
Error
BLOCK IS TOO HOT!
The temperature of the heating block is greater than the
selected temperature. Wait until the temperature of the
PLEASE WAIT
heating block decreases.
INIT ERROR
The instrument is defective. Contact Customer Service.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the
appropriate distributor or refer to the company website for contact information.
Replacement parts
Description
Protective cover
Accessories
Description
Reducing adapter, 20 > 16 mm
Reducing adapter, 16 > 13 mm
10 English
Solution
Quantity
Item no.
1
LZT048
Quantity
Item no.
1
HHA155
1
2895805
Replacement parts and accessories (continued)
Description
Tube, 16 x 100 mm
Tube, 16 x 100 mm
Thermometer, 0 to 200 °C
Quantity
Item no.
1
2275800
6
2275806
1
4565500