Temperature Dropping
Check the Dongle
Set Point Exceeds Sample
Boiling Point
Temperature Unsteady

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Heat deflector shield
Power cord, North America, 230 V, 15 A with C19 connector
Test tube and cooling rack, 40-position
Cold finger set
Digestion tube set, 25 mL and 50 mL, round bottom
Replacement dongle for controller
Fuse, 15 A, 3 AB
Accessories
Description
Calibrated test tube, 25 mL and 50 mL, round bottom
Cold finger, for calibrated digestion tubes
Boiling chips, 250 g
Table 5 Safety messages (continued)
Possible cause
The sample may have evaporated.
An endothermic reaction has
occurred.
The dongle is not connected.
The temperature set point is higher
than the sample boiling point.
The rate of evaporation of the sample
is too high.
The sample has evaporated.
Solution
Make sure that there is sample in the
digestion tubes.
Make sure that the method in the controller
matches the written method. Repeat the
test.
Put the dongle in the COM port.
Change the temperature set point.
Change the temperature set point.
Make sure that there is sample in the
digestion tubes.
Quantity
2/package
each
each
40/package
40/package
each
2/package
Item no.
18091L
18092L
18093L
18032L
18094L
18095L
18105L
Item no.
18090L
18032
18501
English 19