Télécharger Imprimer la page

Hach EZ Serie Manuel De L'utilisateur page 28

Masquer les pouces Voir aussi pour EZ Serie:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
9.4 Prepare and replace the reagents
Fire hazard. The user is responsible to make sure that sufficient precautions are taken
when the equipment is used with methods that use flammable liquids. Make sure to obey
correct user precautions and safety protocols. This includes, but is not limited to, spill
and leak controls, proper ventilation, no unattended use, and that the instrument is never
left unattended while power is applied.
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Do not mix new reagents with old reagents. Discard the old reagents before new reagents are
added to bottles.
1. Reagents and solutions are supplied by the user. Use only reagents supplied by a certified
company. As an alternative, follow the instructions from the Method&Reagent Sheet from the
specific application (EZxxxx) on the manufacturer's website to prepare the reagents.
2. Discard the old reagents from the bottles. If necessary, rinse the bottles with tap water.
3. Fill the bottles with new reagents. Make sure that the tube touches the bottom of the bottle. Make
sure that the tube is not twisted and does not have a blockage.
9.5 Examine and clean the electrode
The electrode maintenance is based on the type of electrode. Refer to the information supplied with
the electrode.
9.6 Calibrate the pH electrode
The calibration procedure is based on the type of electrode. Refer to the information supplied with
the electrode.
9.7 Calibrate the analyzer
The calibration procedure of the analyzer is based on the analyzer method. Refer to
9.8 Clean the analyzer components
Do a cleaning cycle to automatically clean the analyzer components. Refer to
If the cleaning cycle does not remove all of the dirt in the analyzer components or unclog the tubing
and valves, do a manual cleaning as follows:
1. Use a syringe filled with demineralized water to flush tubings, pumps and valves to remove
blockages.
Replace the tubing and valves that stay clogged.
Note: If micropumps stay clogged, examine the micropump duckbills and replace if necessary. Refer to
Replace the micropump duckbills
2. Drain and disassemble the analysis vessel. Clean the analysis vessel components with a moist
cloth. Dry with a soft cloth. Refer to
3. Make sure that all tubing connected to the analysis vessel is at the correct position after
maintenance.
28 English
W A R N I N G
C A U T I O N
N O T I C E
on page 33.
Figure
10.
Do a
calibration.
Do a clean
cycle.

Publicité

loading