G Connecting The Suction Pump; H Arrangement Of Isosun Lamp For Isotron, Isoplus, Isoplus Lite - eurodent ISOTRON Notice D'installation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ISOTRON - ISOPLUS - ISOPLUS LITE - ISOCONCEPT
G - CONNECTING THE SUCTION PUMP
H - ARRANGEMENT OF ISOSUN LAMP FOR ISOTRON, ISOPLUS, ISOPLUS LITE
The dental unit shall be set according to the type of Isosun lamp installed, i.e. with photo cell or without photo cell. In order
to carry out such programming, switch on the dental unit by keeping the keys
one of the two following arrangements:
control for Isosun lamp without photo cell
control for Isosun lamp with photo cell
By keeping the keys
or the key
to store
Release the keys to confirm the selected arrangement and to restore the usual operating order.
46
6 7
6 7
and
pressed, press the key
(Isosun without photo cell).
When picking up any of the two suction tubings, the dental unit
provides a contact between the two white wires from the black
cable (i.e. a switch which conducts when a tubing is up and cuts
off when tubings are at rest).
Therefore, the pump must be provided with control switchboard.
In case of pumps as Dürr or Cattani equipped with control
switchboard, which require a simple switch between two preset
contacts (generally 1 and 3), connect such contacts to the two
brown ones 6 and 7 in front of the white wires, as the figure
shows.
G1
to set the
and
pressed: the display will show
arrangement (Isosun with photo cell),
Installation instructions

Publicité

Table des Matières
loading

Produits Connexes pour eurodent ISOTRON

Ce manuel est également adapté pour:

IsoplusIsoplus liteIsoconcept

Table des Matières