Electric connections in the connecting box: the box receives several cables: the black cable of the chair foot-control (male flat
plug connected to the black cable for chair control), the black cable for chair movement control (female flat plug connected
to the chair foot-control black cable), the black four-wire cable for serial connection to the dental unit (not to be used), and
the white three-wire cable (brown, blue, yellow-green) for chair mains supply.
Coming from the dental unit:
one black three-wire mains supply cable (blue, brown, yellow-green), one white three-wire cable (blue, brown, yellow-green)
another black cable with two black and two white wires, one black cable with nine-wire female flat plug for parallel connection
with the chair, one black cable of the dental unit foot-control and one black cable with two fast-on plugs. There are four sockets
on the bridge-p.c.b., while the safety connectors (a black and a white one) have ten terminals.
seven-wire parallel
cable from chair
foot-control cable of dental unit
cable
from chair
foot-control
nine-wire parallel
cable from dental unit
white wires
white cable to dental unit
black wires
black cable
white cable to chair
220 VAC power mains
P.
C
F3
Plug both fast-on wires to the main solenoid valve for air and water (24V).
The dental unit foot-control cable and the nine-wire parallel cable must be plugged into the relevant connectors. Disconnect
the seven-wire parallel cable of the chair from the one of the chair foot-control and plug them in the relevant connectors of
the bridge-p.c.b. The bridge-p.c.b. must not be used if the dental unit is not there or it consists of a simple Isoconcept cuspidor;
in this case leave the seven-wire parallel cable of the chair and the one of the chair foot-control one plugged into the other.
The white wires (for closing the suction pump control circuit) must be connected to the first two brown terminals, while the black
wires must be connected to the other two (to close the circuit for nurse bell or any other service circuit).