Insert the lamp-stand into its housing and lock it by tightening
the two screws already fitted on the lamp (body inner side).
Make a test chain run through the lamp-stand and bind it with
adhesive tape to the cable coming out the pin of the lamp arm.
By pulling the test chain, insert the pin into the stand, then
connect the lamp cable to the 3-4 terminals of the M4 socket on
the AT/BT rod (on the chair side).
(For the ISOCONCEPT model, the terminal in question is the XT1
on the body p.c.b.).
E12
Position the bowl in its housing.
E13
Connecting box: fix the solenoid valve unit for air and water on
the left side, by using the two screws supplied.
Pay attention to the flow direction inside the valves (shown by
an arrow on the brass body) when joining the air and water
hoses. The electrical connectors unit shall be fastened on the box
right side, by means of the screws and nuts provided.