In the same way, the three wires of the white thick cable shall be connected to the terminals of the white connector (blue to N terminal,
brown to L terminal and yellow/green to earth terminal). To the terminals mentioned the wires of the white cable for feeding the chair
shall be connected according to their colours and L, N and earth signs. Leave the black cable from the chair footcontrol connected
to the black cable for chair movements control. Finally, the wires of the black feeding cable coming from the dental unit shall be
connected to the three terminals of the black connector (blue to N, brown to L and yellow/green to earth). To the terminals mentioned
shall be connected the wires of the cable coming from power mains (220 VAC). Take care to protect all of the hanging connections
by the plastic shield supplied among the accessories.
Installation instructions
ISOTRON - ISOPLUS - ISOPLUS LITE - ISOCONCEPT
suction
dental unit
RL
Bianco White Blanc Weiß Blanco Nero Black Noir Schwarz Negro
N
B 4 S 3 T 2
L 1
Aspirazione
Suction
Aspiration
Absaugung
Aspiración
220 V
24 V max
1 A max
white cable
pump
for chair
gear-case
F5
assistant
dental unit
call RL
T 1
N
L 1
220 V
24 V max
1 A max
power
bell circuit
mains
Connect the foot-control hoses colour-to-colour by the relevant
couplings, keeping the correspondences according to the
colours of the hoses.
Replace the parallelogram lid and fasten it by means of the two
brass nuts/tie rods, then close the connecting box by using the
proper guard. Mount the two lateral panels of the dental unit
body and fasten the cord provided with a cable terminal to the
screw placed on the external panel.
F4
chair earth
45