SERVICE REQUIREMENT

The unit to be installed is CLASS
with electronic bistoury or electrobistoury with attached part of TYPE BF
0051
The apparatus comply with the fundamental requirements of EC Directive 93/42.
Classification headings (EN60 601-1):
5.3
Common equipments (equipments with no waterproof casing);
5.5
Equipments which cannot be used in presence of an anaesthetic mixture inflammable by air or by oxygen or by nitrous
oxide;
5.6
Equipments for continuous working with intermittent load.
Protection earth
Warning, see enclosed documents
Power:
Operating system:
Max. lifting capacity:
The floor must be in accordance with DIN 1055 B1.3 and DIN 18560 T1 regulations.
It is necessary to instal an outer bipolar main switch to wall, in compliance with the law and with the following electric
characteristics: 250 V / 10 A for disconnecting the unit from the main feed in case of intervention or inspection in it.
The electrical system must be in accordance with IEC regulations 64.4, and made up of difference protective conductors (with
index < 30 mA) at 16A/250V.
Moreover, earthing must conform to the art. 3.2.01 of the regulations mentioned above.
Water pressure at the unit inlet must range from 1,8 to 3,0 bar. The pipes must be capable to supply at least 5 l/minute of
water flow rate.
Water characteristics must conform to DIN 1988 rules, with a hardness gradient ≤ 8° dH (corresponding to 1,432 mmol/lit).
We recommend installing a pressure reducing tap at the water hose outlet (preferably of 15 mm Ø).
The air pressure provided by the compressor must range from 5,0 to 7,0 bar. Use a dry compressor, with air output of 60
lit/min. at 5 bar. We recommend installing a tap at the hose outlet (preferably of 15 mm Ø).
The diameter of the emptying hose must be 40 mm, and 32 mm at its inlet; moreover the descent gradient must be at least
1 cm/mt, and the bend angle should not be less than 135°.
The diameter of the suction pipe must be approx. 32/40 mm. We recommend following the instructions provided by the suction
pump's manufacturer.
All arrangements must be carried out according to the indications provided on the following page and consulting the full scale
installation plan (code 633580010) attached to each dental unit.
Please refer to Spare Parts catalogue and apply to EURODENT for further information about what is not included in it.
EURODENT accepts responsibility for the equipment safety and reliability if and only if:
- installation, assembly, extension, adjustment and repair operations have been carried out by authorized people.
- the general and electric installation of the room complies with the above written requirements.
- the equipment has been installed in compliance with the "INSTALLATION INSTRUCTIONS".
- the equipment is used in compliance with the "OPERATOR'S HANDBOOK".
26
, with attached parts of TYPE B
I
220 V / 50 Hz / 1350 VA (550 VA only unit) for ISOTRON, ISOPLUS, ISOPLUS LITE.
220 V / 50 Hz / 1000 VA (200 VA only unit) for ISOCONCEPT.
continuous, with intermitting load
120 kg (dental unit and lamp excluded)
,
(ISOTRON, ISOPLUS, ISOPLUS LITE only)
Installation instructions