Into the shaped hole on the parallelogram arm of the chair first
insert the suction hose, then the electric cables wrapped with
tape to the Rilsan hoses (green and white), finally insert the
drain hose. Lead all hoses and cables up to the connecting box.
E3
ISOTRON, ISOPLUS, ISOPLUS LITE
The accessories boxes contain a foot-control, the suction
tubings with their proper hoses, an assistant syringe (on
request) with a clip, four screws, a grey cord to connect the foot-
control cable to the connecting box, main solenoid valves for
water and air and safety electrical connectors.
The cord has a squared-plug terminal and a circular-base
terminal: push the first terminal mentioned into the hose which,
inside the base footboard, joints the outside to the connecting
box. Then, pull it until the frame of the circular base adheres to
the frame of the hose. Tighten the four self-tapping screws in the
holes of the frame, so that they penetrate completely the
polyurethane and fasten properly the cord base.
E4
ISOCONCEPT
The accessories boxes contain the foot-control, the suction tips
with their proper hoses, the nurse syringe (on request), four
screws, 2 couplings, the water and air main valve, safety
electric connectors and bridge-p.c.b.
Slip electric cables and hoses of the foot-control through the
sleeve which, inside the base footboard, leads to the connection
box, and push the flange of the foot-control cord against the
sleeve frame. Tighten the four small self-tapping screws in the
holes of the frame, until they have entered completely the
polyurethane and lock the flange.
E5