E7
Pump excess voltage.
Pump excess current.
E8
Foot Control
E9
abnormality.
LED excess voltage
E10
(Optic Motor).
LED excess current
E11
(Optic Motor).
Error code
USB flash drive other than the type stipulated is connected
UE0
("EO" will be displayed on the SPEED screen).
UE1
USB flash drive module circuit error.
Data cannot been stored on USB flash drive for the one of
the following reasons:
• Lack of available memory on USB flash drive.
UE2
• Use of a USB flash drive on which copying is prohibited.
• File name is already in use.
• Other USB flash drive related error.
Error
Screen is blank.
Rotation number display is "--" and
numbers are displayed and numbers
alternate.
Will not rotate even if the Foot Control
is pressed.
The Irrigation tube is incorrectly
fitted and loading the pump roller.
The pump fails.
Connection failure of the connector.
Failure of the inside of the Foot
Control.
Failure of LED circuit.
Check if the saline solution hasn't
penetrated the motor lamp.
Failure of LED circuit.
Cause of Error
Cause of Error
Plug to power source is not properly
inserted.
Blown fuse.
Malfunction of power switch.
Power is on while Foot Control is being
pressed.
Foot Control malfunction.
Control Unit malfunction.
Cord connection is loose.
Malfunction of Control Unit or Foot
Control.
17
Check the Irrigation tube.
If an Error Code is displayed even when the
Irrigation tube is normal, request repair.
Check the connection state of the connector.
The Error Code is displayed even when the
connector is normally connected, request
repair.
Request repair.
Remove the saline solution completely.
Request repair.
Remedy
Please connect using the proper type of
USB flash drive.
Notify retailer where device was purchased.
• Replace USB flash drive.
• Remove lock that prevents copying.
• Change the file name.
• Read instructions for USB flash drive use.
Remedy
Check to make plug is in.
Replace fuse.
Request repair.
Remove foot from Foot Control.
If the problem is solved by removing
the Foot Control, the Foot Control has
malfunctioned. Repair is required.
If the problem is not solved by
removing the Foot Control, the unit
has malfunctioned. Repair is required.
Check connection.
If the problem is not solved by
checking the connection, system
repair is required.

Publicité