NSK Surgic Pro Mode D'emploi page 23

Masquer les pouces Voir aussi pour Surgic Pro:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Guidance and manufacturer's declaration - electromagnetic immunity
Surgic Pro series is intended for use in the electromagnetic environment specified below.
The customer or the user of the Surgic Pro series should assure that it is used in such an environment.
Immunity test
IEC/EN60601 test level
Conducted
3V rms
RF IEC61000-4-6/EN61000-4-6
150 kHz to 80MHz
Radiated RF
3V/m
IEC61000-4-3/EN61000-4-3
80MHz to 2.5 GHz
NOTE1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the Surgic Pro series is used exceeds the applicable RF
compliance level stated above, the Surgic Pro series should be observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the Surgic Pro series.
b Over the 150kHz to 80MHz frequency range, the field strength should be less than 3V/m.
Cables and accessories
Maximum length
Micromotor with Motor Cord
2.0M
Foot Controller with Cord
2.0M
AC Power Cord
2.0M
Recommended separation distances between portable and mobile RF communications equipment and the Surgic Pro series.
The Surgic Pro series is intended for use in an electromagnetic environment in which radiated RF disturbances are control LCD. The customer or the user of the
Surgic Pro series can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the Surgic Pro series as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance 'd' in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where 'P' is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Compliance level
3V rms
3V rms
Shield
Complies with
Unshielded
RF emissions, CISPR11:
Unshielded
Harmonic emissions:
Unshielded
Voltage fluctuations/ flicker emission:
Electrostatic discharge (ESD):
Electrical fast transient/burst
Surge:
Voltage dips, short interruptions and voltage variations on power supply input lines:
Power frequency(50/60Hz) magnetic field:
Conducted RF:
Radiated RF:
Separation distance according to frequency of transmitter
150kHz to 80MHz
d =1.2
P
0.12
0.38
1.2
3.8
12
22
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the Surgic Pro series, including
cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d
= 1.2
P
d
= 1.2
80MHz to 800MHz
P
d
= 2.3
800MHz to 2.5GHz
P
Where is the maximum output power rating of the transmitter in
P
watts (W) according to the transmitter manufacturer, and ( ) is
the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters as determined by an
electromagnetic site survey(a) should be less than the compliance
level in each frequency range(b).
Interference may occur in the vicinity of equipment marked with
the following symbol:
m
80MHz to 800MHz
d =1.2
P
0.12
0.38
1.2
3.8
12
d
Class B/ Group 1
IEC61000-3-2/EN61000-3-2
IEC61000-3-3/EN61000-3-3
IEC61000-4-2/EN61000-4-2
IEC61000-4-4/EN61000-4-4
IEC61000-4-5/EN61000-4-5
IEC61000-4-11/EN61000-4-11
IEC61000-4-8/EN61000-4-8
IEC61000-4-6/EN61000-4-6
IEC61000-4-3/EN61000-4-3
800MHz to 2.5GHz
d =2.3
P
0.23
0.73
2.3
7.3
23

Publicité

Table des Matières
loading

Table des Matières