General Information - Philips InnoSpire Go Mode D'emploi

Masquer les pouces Voir aussi pour InnoSpire Go:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

1. Handset
2. Mouthpiece assembly
3. AC power adapter
4. LiteTouch medium mask (age 2 - 5 years)
5. Mask adapter
6. On/Off button and LED indicator
Instructions for use
Please read these instructions carefully before first use. If you do not understand any portion
of these instructions, contact your healthcare provider or call Philips Customer Service at
1-724-387-4000.

General information

CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician or
licensed healthcare professional.
Intended use: The InnoSpire Go is a vibrating mesh nebulizer system designed to aerosolize
liquid medications for inhalation by the patient. The device may be used with pediatric
(2 years and older), defined by the prescribed medication, and adult patients in the home
environment or in a hospital/clinic setting.
Keep these Instructions for future reference
Retain carton and packing materials for storing the unit or for product returns.
Cautions
It is recommended to have a backup device (e.g. MDI) for respiratory delivery in case
a situation arises when your nebulizer cannot be used.
InnoSpire Go is intended for SINGLE PATIENT USE, to deliver multiple doses.
It is not to be used by patients that are unconscious or not breathing spontaneously.
The patient or the patient's care giver is the intended operator of the device.
Only use this nebulizer with medications prescribed by your physician.
Do not place or store the product where it can fall into water.
Do not submerge the handset in water or other liquid cleaning agents.
If the handset is accidentally dropped into a liquid, immediately disconnect the AC adapter
from the wall outlet, prior to removing the handset from the liquid.
Any liquid spilled on the handset should be allowed to dry before operating.
Do not position the product where it is difficult to disconnect the plug.
Adult supervision is necessary when this product is used by, on, or near children or
physically challenged individuals or individuals with learning difficulties.
4
7. Medication chamber lid
8. Medication chamber
9. Mouthpiece assembly release button
10. Power socket (back view)
11. Carry case (not shown)
Use this product only for its intended use as described in this manual. Failure to do so
may compromise performance. Only use accessories provided with the nebulizer and
recommended by the manufacturer. Store in a clean place out of the reach of children.
Do not disassemble or modify the nebulizer in anyway. There are no serviceable parts.
The battery is non-replaceable.
The battery has been fitted at the date indicated on the unit carton and has a shelf life
of 18 months past this date.
Never charge or operate this product if it has a damaged cord or plug, if it is not working
properly, if it has been dropped or damaged, or dropped into water.
Do not connect to other equipment not described in these instructions.
Do not charge your device on an aircraft.
Keep the power cord away from heated surfaces.
The nebulizer must be operated using the specified power sources.
Always check the inside of the mouthpiece assembly or mask for any debris before use.
Do not use while operating a vehicle.
Never use when lying down.
When operated at an ambient temperature 40˚C/104˚F the handset can reach temperatures of
up to 43˚C/109˚F. The device should not be used for more than 10 minutes during this scenario.
Do not use in an anesthetic or ventilator breathing system.
InnoSpire Go and all its parts (including battery) must be disposed of properly and according
to the local regulations in force (for example the WEEE directive).
Small parts can be inhaled or swallowed. In addition, the cable, due to its length, may result
in provoking strangulation or asphyxiation. Do not leave the device alone with a small child
or physically challenged individual or individual with learning difficulties.
The nebulizer should not be used around flammable substances e.g. oxygen, nitrous oxide,
or in the presence of a flammable anesthetic mixture.
Never poke the mesh, or clean it with any sharp objects. This could damage the mesh and
prevent your device from operating properly.
Do not autoclave the device.
Report unexpected operation or events to Philips.
Precautions are to be taken in the event of changes in the performance of the device, please
refer to the troubleshooting section.
How to use your InnoSpire Go
After unpacking the nebulizer check you have all the items listed and that there is no visible
damage or defects. Contact your product distributor or Philips customer service if anything
is missing or damaged. Prior to first use clean the mouthpiece assembly as per the cleaning
instructions and fully charge the battery. Ensure the device is disconnected from the power
during assembly and disassembly.
ENGLISH
5

Publicité

Table des Matières
loading

Table des Matières