A: Cycle selection button
B: Ext w-o (or Spray) button for selection of the model of the handpiece(s)
C: Instruments selection and Oil volume button (Short or Long mode)
D: Start-up button. This button can also be used to stop the unit during cycle (with a long press).
E: Cycle progress operating light indicators (LEDs)
F: Warning operating light indicators (LEDs) for consumables state and for drawer, air, door state
1, 2, 3, 4: Indicator of the type of instrument (if the indicator is lit: instrument with external spray or without spray)

3.5.2. Connecting the instruments

The instruments have to been connected in accordance to the description done in

3.5.3. Selecting the program

The treatment of the instruments is achieved by a cycle composed of 3 different steps:
1 cleaning phase : 7 min (for 4 instruments)
1 disinfection phase : 7 min (for 4 instruments)
1 lubrication phase : 1.5 min (for 4 instruments)
The "CYCLE" button enables the operator to choose between the 3 maintenance cycles:
Full maintenance cycle: Cleaning/Disinfection/Lubrication
Partial maintenance cycle: Cleaning/Lubrication
Short maintenance cycle: Lubrication only (similar function as for Care3plus)
Full mode
th
A 4
pressure on the Cycle button manages the user to reset the choice of the cycle.
iCare+ User Manual V11.4 • 09/2017
À: Operating light indicators (LEDs) for cycle selected
Partial mode
Figure 20: Selection of the type of program for iCare+ device
chapter 3.4.
Short mode
p. 25