• The disposable handpiece and handpiece sleeve are consumable items. Dispose of each as
medical waste after every use.
CAUTION
• Make sure to install a handpiece sleeve before every use to prevent contamination of the
motor handpiece.
4
Preoperational checks
Follow the check procedure below before use. If any abnormalities are found, stop using the handpiece
immediately and contact your Authorized NSK Dealer.
1) Check the head of iProphy mobile is firmly fixed.
2) Operate the product outside the patient's mouth and check the disposable handpiece head rotates properly.
During rotation, check for abnormalities such as abnormal rotation, vibration, noise.
5
Operation
5-1 Charging the Battery
1) Insert the AC adaptor plug into a commercial power supply.
2) Insert the motor handpiece into the stand. The battery charge indicator lamp lights up and charging starts.
3) When the battery charge indicator lamp turns off charging is completed.
The
lamp lights up when remaining battery power becomes low. When the
immediately charge the battery.
• Do not insert the plug into any outlet other than the one for commercial power supply.
• Turn off the motor handpiece before charging.
CAUTION
• The standard charging time is Approx. 90 minutes. This may vary depending on the battery
condition, age, temperature and other factors. With older batteries the charging time and
operating time may become significantly shorter.
• Only use the supplied AC adaptor.
• Gently place the motor handpiece into the stand. A strong impact may cause handpiece
failure.
• When placing the motor handpiece into the stand pay attention to the insertion direction.
Excessive force is not required and it may cause damage to the handpiece. The handpiece
may also be damaged if dropped.
• When charging the battery, be sure to remove the handpiece sleeve from the motor
handpiece.
• The stand is exclusively for "iProphy mobile" motor handpiece. Never use the stand for any
other purpose.
• The battery will get warm during charging, but this is not a failure. Inserting or removing the
plug at short intervals (approx. 5 minutes) may generate heat and interfere with charging.
Charge the rechargeable battery for at least 10 minutes.
• If not used for a long period charge the battery prior to use. If not charged periodically the
battery will become fully discharged and will not charge properly. If this occurs contact your
authorized NSK Dealer to purchase a new battery.
• Do not place into the stand any object (wire, safety pin, etc.) other than the motor handpiece
of this product. This may cause burns or a failure due to overheating.
8
lamp is illuminated