Disinfection Instructions - Invacare Ocean Ergo Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 65
Cleaning the Toilet Pan (optional)
The toilet pan can be cleaned in an autoclave at following
temperatures/times:
90 °C for 1 minute
85 °C for 3 minutes
80 °C for 10 minutes.

8.2.4 Disinfection Instructions

Information on recommended disinfectants
and methods can be found on https://vah-
online.de/en/for-users.
In Domestic Care
Method: Follow the application notes for the used
disinfectant and wipe-disinfect all accessible surfaces.
Disinfectant: Ordinary household disinfectant.
Drying: Allow the product to air-dry.
In Clinical and Long Term Care
Follow your in-house disinfection procedures and only use
the disinfectants and methods specified therein.
1639501-C
Maintenance