Chapter 14 Cleaning and Disinfection

14.1 Cleaning the Device and Accessories

Your device should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place, the
device should be cleaned more frequently. Before cleaning the equipment, consult your hospital's regulations for
cleaning the device.
Recommended cleaning agents are:
Sodium hypochlorite bleach (diluted)
Hydrogen peroxide (3%)
75% Ethanol
70% Isopropanol
To clean your device, follow these rules:
Switch off the monitor and disconnect the power cable before cleaning.
Kept the monitor from dust.
It is recommended to clean the outer shell and screen of the monitor to keep it clean. Only non-corrosive
cleanser such as clear water is permitted.
Wipe the surface of the monitor and transducers with an Ethanol impregnated wipe, and dry it with dry and
clean wipe or simply air-dry.
This monitor can be disinfected, please clean the monitor firstly.
Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage.
Clean the exterior of the connector only.
Dilute the cleanser.
Do not use scrub materials.
Do not let any liquid flow into the shell or any parts of the monitor.
Do not let the cleanser and disinfectant stay on its surface.
Do not perform high pressure sterilization to the monitor.
Do not put any parts of the monitor or its accessories in the liquid.
Do not pour the disinfectant on its surface while disinfectant.
If the monitor is accidentally wet, it should be thoroughly dried before use. The rear cover can be removed
by qualified service technician to verify absence of water.
Never use this machine in an environment with inflammable gas.
Avoid being hit by lightning. The power cable should be plugged into an outlet with grounding wire. Do not
Operator's Manual for Patient Monitor