Colour temperature of LEDs
Expected service life of the device
Maximum radiance output of the TL 45 Perfect Day
Radiance output
EUVA: Eye UV-A
ES: Actinic UV skin & eye
EIR: Infrared radiation
hazard exposure limits
for the eyes
EH: Thermal hazard for
the skin
LB: Blue light
LR: Retinal thermal
Subject to technical changes.
The serial number is located on the device.
Brightness: 10,000 lux (this point about the light intensity is merely for the purpose of information. With regard to
the standard IEC 60601-2-57, this light source is classified as an Exempt Group)
Notes on electromagnetic compatibility
The device complies with the EU Medical Devices Directive 93/42/EEC (therapy setting), the German Medical
Devices Act (Medizinproduktgesetz) and the European standard EN 60601-1-2 (in accordance with CISPR-11,
IEC61000-3-2, IEC61000-3-3, IEC 61000-4-2, IEC61000-4-3, IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6,
IEC 61000-4-8, IEC 61000-4-11) and is subject to particular precautions with regard to electromagnetic
compatibility.
• The device is suitable for use in all environments listed in these instructions for use, including domestic envi-
ronments.
• The use of the device may be limited in the presence of electromagnetic disturbances. This could result in
issues such as error messages or the failure of the display/device.
• Avoid using this device directly next to other devices or stacked on top of other devices, as this could lead to
faulty operation. If, however, it is necessary to use the device in the manner stated, this device as well as the
other devices must be monitored to ensure they are working properly.
• The use of accessories other than those specified or provided by the manufacturer of this device can lead to
an increase in electromagnetic emissions or a decrease in the device's electromagnetic immunity; this can
result in faulty operation.
• Failure to comply with the above can impair the performance of the device.

13. Mains part

Model no.
Input
Output
therapy
active
relax
10,000 hours
Risk group
classified in
accordance with
IEC 60601-2-57:
2011
Exempt Group
4.784x10
Exempt Group
7.249x10
Exempt Group
1.045x10
Exempt Group
2.487 e
Exempt Group
3.094 x10 W·m
Exempt Group
3.918x10
MP 45-240150-AG MP 45-240150-AB
100–240 V ~ 50/60 Hz
24 V DC, 1.5 A, only in conjunction with GCE503
6,500 K ± 300 K
4,300 K ± 300 K
3,000 K ± 300 K
Maximum value
Therapy
-5
W·m
-2
7.349x10
-4
W·m
-2
2.205x10
-1
W·m
-2
6.252x10
1
W·m
-2
·sr
-1
1.477 e
-2
-1
·sr
7.266 W·m
2
W·m
-2
·sr
-1
9.968x10 W·m
21
Active
Relax
-6
W·m
-2
0 W·m
-2
-10
W·m
-2
0 W·m
-2
-2
7.208x10
1
W·m
-2
·sr
-1
6.182 e
0
-2
-1
·sr
4.15x10 W·m
-2
·sr
-1
7.563x10 W·m
-2
W·m
-2
W·m
-2
·sr
-1
-2
-1
·sr
-2
·sr
-1

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Perfect day tl 45

Table des Matières