Disposal; Specifications - Beurer BM 55 Mode D'emploi

Masquer les pouces Voir aussi pour BM 55:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 39
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical specifica-
tions", the technical alarm will appear on the display indicating
either "     " or " Lo Lo    " . In such cases, you should seek medical as-
sistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.

11. Disposal

Repairing and disposing of the device
• Do not repair or adjust the device yourself. Proper operation
can no longer be guaranteed in this case.
• Do not open the device. Failure to comply will invalidate the
warranty.
• Repairs must only be carried out by Customer Services or
authorised retailers. Before making a claim, please check the
batteries first and replace them if necessary.
• For environmental reasons, do not dispose of the
device in household waste at the end of its service life.
Dispose of the device at a suitable local collection or
recycling point in your country. Dispose of the device in
accordance with EC Directive – WEEE (Waste Electrical and
Electronic Equipment). If you have any questions, please
contact the local authorities responsible for waste disposal.
Disposing of the batteries
• The empty, completely flat batteries must be disposed of
through specially designated collection boxes, recycling
points or electronics retailers. You are legally required to
dispose of the batteries.
• The codes below are printed on batteries
containing harmful substances:
Pb = Battery contains lead,
Cd = Battery contains cadmium,
Hg = Battery contains mercury.

12. Specifications

Model no.
Type
Measurement
method
Measurement range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
35
BM 55
M1002
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Cuff pressure 0 – 300 mmHg,
systolic 60– 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 186 mm x W 95 mm x H 56 mm
Approx. 467 g (without batteries)

Publicité

Table des Matières
loading

Table des Matières