Setup
CAUTION
Only use nasal cannula with Lumis HFT, do not use full face masks.
Do not overfill the humidifier as water may enter the device and air tubing.
To avoid heated water flowing down the tube to the patient:
o
place the device lower than the level of the patient's head;
o
ensure the device is placed on a stable surface.
1. With the device on a stable level surface below head height, connect one end of the power cord
into the power supply unit and the other end into the power outlet.
2. Fit the tubing wrap around the air tubing.
See the tubing wrap user guide for detailed information. Humidification performance will be
reduced if the tubing wrap is not fitted.
3. Connect the air tubing firmly to the air outlet located on the rear of the device.
4. Open the humidifier and fill it with water up to the maximum water level mark.
Do not fill the humidifier with hot water.
5. Close the humidifier and insert it into the side of the device.
4