• Do not perform any maintenance tasks while the device is in operation.
• The device should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the device should be observed to verify normal operation in the
configuration in which it will be used.
• The use of accessories other than those specified for the device is not recommended. They
may result in increased emissions or decreased immunity of the device.
• The device has not been tested or certified for use in the vicinity of X-ray, CT or MRI
equipment. Do not bring the device within 4 m of X-ray or CT equipment. Never bring the
device into an MR environment.
• Therapy settings should not be changed remotely for patients in a hospital setting.
• Do not use the device outside its approved operating conditions. Using the device above
an altitude of 2,591m and/or outside the temperature range of 18°C to 28°C, may reduce the
effectiveness of treatment and/or damage the device.
• Connecting the device to the battery of a battery-powered wheelchair may affect the device
performance and may result in patient harm.
• Use only water-based lotions or salves that are oxygen-compatible before and during
oxygen therapy. Do not use petroleum-based or oil-based lotions or salves to avoid the risk
of fire and burns.
• Do not lubricate fittings, connections, tubing, or other accessories of the equipment to
avoid the risk of fire and burns.
CAUTION
• Use only ResMed (or ResMed recommended) parts and accessories with the device. Parts
not recommended by ResMed may reduce the effectiveness of the treatment and/or
damage the device.
• Use only nasal cannulas recommended by ResMed or by the prescribing doctor with this
device.
• Be careful not to place the device where it can be bumped or where someone is likely to
trip over the power cord.
• Keep the area around the device dry, clean and clear of anything (eg, clothes or bedding)
that could block the air tubing, inlet, or cover the device and/or power supply unit while in
operation, as this could lead to overheating of the device.
• Do not place the device on its side as water might get into the device.
• Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or
antibacterial soaps or scented oils to clean the device, the humidifier or air tubing. These
solutions may cause damage or affect the humidifier performance and reduce the life of
the products. Exposure to smoke, including cigarette, cigar or pipe smoke, as well as ozone
or other gases, may damage the device. Damage caused by any of the foregoing will not
be covered by ResMed's limited warranty.
• Leave the humidifier to cool for ten minutes before handling to allow the water to cool and
to make sure that the humidifier is not too hot to touch.
• Make sure that the humidifier is empty before transporting the device.
• Avoid placing the device directly on varnished surfaces during operation, as heat build-up
may damage the finish.
Notes:
The device is not intended to be operated by persons (including children) with reduced physical,
sensory or mental capabilities without adequate supervision by a person responsible for the
patient's safety.
For any serious incidents that occur in relation to this device, these should be reported to
ResMed and the competent authority in your country.