Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 58
Pneumatic flow path
Design life
Device, power supply unit:
Cleanable humidifier tub:
Air tubing:
General
The patient is an intended operator.
Operator position
The device is designed to be operated within arm's length. An operator should position their line of sight within an angle of
30 degrees from a plane perpendicular to the screen.
Air tubing
Air tubing
Material
ClimateLineAir
Flexible plastic and electrical components
Heated air tubing temperature cut-out: ≤ 43°C
Notes:
• The manufacturer reserves the right to change these specifications without notice.
• The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the nasal cannula.
• Do not use electrically conductive or antistatic air tubing.
• The temperature and relative humidity settings displayed are not measured values.

Displayed values

Value
Flow
1
Results are expressed as STPD (Standard Temperature and Pressure, Dry). (101.3kPa at an operating temperature of 20°C, dry). Use the
table below to convert the STPD flow setting to BTPS (Body Temperature and Pressure, Saturated) flow.
BTPS Flow (L/min)
Altitude (m)
0
500
1000
1500
2000
2500
Measurement system uncertainties
In accordance with ISO 80601-2-74:2017 the measurement uncertainty of the manufacturer's test equipment for measures
of flow is ± 1.5 L/min or ± 2.7% of reading (whichever is greater)
22
1. Flow sensor
2. Blower
3. Pressure sensor
4. Nasal cannula
5. Air tubing
6. Humidifier
7. Device
8. Inlet filter
9. Low flow oxygen input (up to 15 L/min)
3 years
3 months
6 months
Accuracy
1
±6 L/min at 0 to 40 L/min positive flow
Target Flow Setting (L/min)
15
20
17
23
18
24
19
26
20
27
22
29
23
31
Length
2 m
25
30
28
34
30
36
32
38
34
41
36
43
38
46
Inner diameter
15 mm
35
40
39
45
42
48
45
51
48
54
51
58
54
62

Publicité

Table des Matières
loading

Table des Matières