Table
6.1 Warning screens (continued)
Screen
Rechargeable
battery only
Rechargeable
battery is in a
discharged state
 Warning: If the neurostimulator battery is
allowed to overdischarge, you may not be
able to recharge it. However, your doctor
may be able to restore the battery
function.
Refer to "The rechargeable neurostimulator
battery" on page 100 for more information.

Communication screen

The Communication screen appears when
the patient programmer is trying to
communicate with your neurostimulator. If
there is a problem, the displayed screen
Cause and action
The rechargeable
neurostimulator battery is in a
discharged state and therapy
has stopped.
Charge the neurostimulator
battery now. Refer to the
manual packaged with the
charging system.