Battery function is restored; however,
charging sessions may be more
frequent because battery capacity has
been reduced.
Battery function is not restored and the
neurostimulator must be surgically
replaced. Battery function is not
restored because:
the neurostimulator battery is
permanently damaged.
the neurostimulator battery has
been overdischarged and restored
twice before. The third time the
battery is overdischarged, the
neurostimulator will reach end of
service. Surgery is required to
replace the neurostimulator.

Rechargeable battery overdischarge

You can avoid the consequences of an
overdischarged battery and the loss of
therapy by keeping the battery charged.
37642 2015-04  English 
107