Installing the Sensor

1. Take either sensor and apply a continuous lengthwise 3mm (1/8-inch) bead of coupling
FUG1010 IP65 NEMA 4X Quick Start
Operating Instructions, 06/2014, A5E02951527-AC
compound across the center of the sensor emitting surface.
F-Connector
Angled Edge
Sensor
Figure 3-8
Sensor Callouts
Front View
Spring Clip (Not present on some
models)
Sensor Clamping Screw
Spacer Bar
Mounting Frame
Figure 3-9
Sensor Installation
3.5 Sensor Installation
Front Face
Emitting Surface
Coupling Compound
Pipe
Mounting Strap
Note: Optional 2nd Mounting Strap
shown. Larger pipes over 76cm (30
inches) may need additional support..
Spacer Bar Platform and Clamping
Screw
Spacer Bar Reference Hole
Orientation for Single Beam Sensor at
9 o'clock position
Orientation for Dual Beam Sensor at
10 & 2 o'clock positions