Connecting
3.2 Transmitter Wiring
Wiring Temperature Sensor Board
1. Using a flat-blade screwdriver, loosen Terminal Block TB1 screw.
2. Wire the RTD 992EC temperature cable as shown in the table below:
Note
*For cathodically protected pipes, do not attach blue #5 wire at RTD end of cable.
3. Replace I/O Board and secure with four captive screws paying careful attention to pin
alignment.
4. Replace Access Cover and finger tighten captive thumbscrew.
Note
TB3 and TB4 are also active analog inputs. See wiring table below.
24
992EC Series Cable
Wire #1 (Black)
Wire #2 (Orange)
Wire #3 (Brown)
Wire #4 (Red)
Wire #5 GND/SHLD (Blue)
Terminal TB1
To TB1--1
To TB1--2
To TB1--3
To TB1--4
*To TB1--5
FUG1010 IP65 NEMA 4X Quick Start
Operating Instructions, 06/2014, A5E02951527-AC

Publicité