Important
The device is not suitable for continuous operation;
after 20 minutes of operation it must be switched off
for 40 minutes.
Note
During treatment, hold the atomiser straight (verti-
cal), otherwise the atomisation does not work and
correct function cannot be guaranteed.
Important
Essential oils, cough medicines, solutions designed
for gargling, and drops for application to the skin
or for use in steam baths are wholly unsuitable for
inhalation using a nebuliser. These substances are
often highly viscous and can impair the correct func-
tion of the device and hence affect the effectiveness
of its application in the long term.
In the case of oversensitivity of the bronchial
system, medicines containing essential oils can
sometimes cause acute bronchiospasm (a sudden,
cramp-like tightening of the bronchi accompanied
by breathlessness). Always ask your doctor or phar-
macist for advice!
6. End of inhalation
If the spray is only output sporadically or if the noise
changes during inhalation, you can end the treat-
ment.
• Switch off the nebuliser after treatment using the on/
off switch [3] and disconnect it from the mains.
After use, replac
e the atomiser [9] in its holder [7].
7. Clean the device
For information about this, see "Cleaning and dis-
infection" on page 15.

7.Changing the filter

In normal operating conditions, the air filter must be
replaced after approx. 500 operating hours or one year.
Please check the air filter regularly (after 10-12 nebu-
lisation procedures).
If the filter is very contaminated or blocked, replace
the used filter. If the filter has become wet, it must
also be replaced with a new one.
Caution
• Do not try to clean and re-use the used filter!
• Only use original filters supplied by the manufacturer.
Otherwise, your nebuliser may become damaged or
effective therapy can no longer be guaranteed.
• Do not repair or maintain the air filter while it is in use.
• Never operate the device without a filter.
To replace the filter, proceed as follows:
Caution
• First switch off the unit and unplug it.
• Allow the unit to cool down.
1.
Remove filter cap [2] upwards.
Note
If the filter remains in the unit when the cap is re-
moved, remove the filter from the unit using twee-
zers or similar.
2. Replace the filter cap [2] containing the new filter.
3. Check that the filter cap is firmly in place.

8. Cleaning and disinfection

Atomiser and accessories
Warning
Observe the following hygiene instructions in order
to avoid any health risks.
The atomiser
[7] and accessories are intended for
multiple use. Please note that cleaning requirements
differ according to the different application areas.
Notes:
• The atomiser and accessories must not be me-
chanically cleaned with brushes or similar as this
can cause irreparable damage and the success of
the treatment can no longer be guaranteed.
• For additional requirements regarding the necessary
hygienic preparations (hand washing, handling of
medicines or inhalation solutions) in high-risk groups
(e.g. cystic fibrosis patients), contact your doctor.
• Ensure that all parts are properly dried following all
cleaning and disinfection. Residual moisture or water
droplets can result in an increased risk of bacterial
growth.
Preparation
• Directly following each treatment, all parts of the
atomiser [7] and any accessories used must be
cleaned of medicine residues and contamination.
• To do this, dismantle the atomiser [7] into its com-
ponent parts.
• Remove the mouthpiece [10] from
the atomiser.
• Dismantle the atomiser by twisting
the upper section anti-clockwise
against the medicine container [9].
• Remove the nozzle attachment [8]
from the medicine container [9].
• The nebuliser is later reassembled
in the reverse order.
15
8
9