Test Frequency
(MHz)
Radiated RF
2450
IEC61000-4-3
(Test speci-
fications for
ENCLOSURE
PORT IMMUNITY
5240
to RF wireless
communications
5240
equipment)
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME
SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT, and using higher
IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Minimum separation distances for higher
IMMUNITY TEST LEVELS shall be calculated using the following equation:
6
E
P
d
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST LEVEL in V/m.
Band a)
Service a)
(MHz)
2 400 – 2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
5 100 – 5 800
WLAN 802.11
a/n
Modulation b)
Modulation b)
(W)
Pulse
2
modulation b)
217 Hz
Pulse
0,2
modulation b)
217 Hz
66
Distance
IMMUNITY TEST LEVEL
(m)
(V/m)
0,3
28
0,3
9