Tecno-gaz Europa B EVO Mode D'emploi page 69

Autoclave
Masquer les pouces Voir aussi pour Europa B EVO:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 83
A 501
Pressure not maintened
A 551
Irregular 1° temperature sensor
A 552
Irregular 2° temperature sensor
A 641
Overly high pressure
A 651
Overlay high temperature
A 661
Irregular temperature comparison
A 701
Working pressure not reached
A 751
Low temperature
A 761
Irregular pressor sensor
A 801
Irregular unloading pressure
Main voltage too low - Temporary lack of
A 901
electric energy
REPLACE RESINS Resins IONIC
SYSTEM exhaust
REPLACE FILTER  Resins OSMOSI SYSTEM
exhaust
The printer's door is not
D
closed
P
Paper is missing
There is low voltage on
Lo
the line
In the event of one of the alarms persisting after some time, consult technical assistance.
Try to close the printer's door
Try to connect the machine to the net electrical worker in
another point; if the problem persists to contact a
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle
• Reset and restart cycle,
• Inform eletric Energy provider
Replacement of resins; refer to
paragraph 16.3
Replacement of filter; refer to
paragraph 16.3
Insert paper
electrician
ENGLISH