Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
The "E" code is relative to errors of the operator, which can be corrected by same operator.
The "A" code is relative to alarms, irregularities of the autoclave, after having carried out the
remedy, if the problem persists it is necessary to request telephonic assistance.
If an alarm message is displayed (CODE "A") the cycle is to be considered
aborted: it will be necessary to repeat all the preparation and sterilization
operations.
To reset alarms and errors, maintain pressed the buttons here indicated until the
intervention of a
BEEP"
"
B-DOOR
ERROR
E 200
Low water level
Low level water cleaned up with loadingfrom
E 250
demineralizer
E 300
Maximum level used water
E 401
Door open
E 451
Door unlocked
High temperature in chamber for vacuum test
E 461
(above 40° C)
A 250
Cycle begins without water
A 001
Interrupted cycle
A 011
Irregular pc display
A 101
Vacuum not reached in 10 minutes
A 111
Vacuum not maintened for first phase VT
A 121
Vaccum not maintened for second phase VT • Reset and restart cycle
Hatch opende during the cycle or problem
A 401
closing
26
MESSAGES OF ERROR OR ALARMS
B-PUMP
+
CAUSE
B-SELECT
+
REMEDY
• Fill tank
• Fill tank
• Empty tank
• Colse door with attention
• Open door and close again
• Shut off and wait 10 minutes
• Control demineralizer device
• Reset and restart cycle
• Shut off and restart autoclave
• Reset and restart cycle
• Reset and restart cycle
• Check the closing system