I-Tech T-ONE REHAB Notice D'utilisation page 147

Masquer les pouces Voir aussi pour T-ONE REHAB:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Medical
Prog
Prog.
Yes/No
U8
Yes
U9
Yes
URO1-2-3 • Stress incontinence (medical program)
Programs suitable for the treatment of stress urinary incontinence in women
and faecal incontinence in men (only U1), designed to strengthen and tone up
the perineal muscles and the muscles of the pelvic floor which have lost force
and contractile capacity, or the sphincter muscles with weak contractile
capacity. The stimulation should be as strong as possible without being
painful. In addition, it helps if the patient performs voluntary contractions
during the stimulation phase. To be associated with appropriate training
exercises for muscle strengthening. Applications: 3-5 sessions per week. Use
the vaginal probe for the treatment of urinary incontinence in women and
the anal probe for fecal incontinence in women and in men.
URO4-5-6 • Urge incontinence (medical program)
Program suitable for the treatment of urge urinary incontinence in women
and faecal incontinence in men (only U4). Low-frequency stimulation which
favours bladder relaxation in case of hyperactivity. The stimulation should be
as strong as possible without being painful. In addition, it helps if the patient
performs voluntary contractions during the stimulation phase. Applications:
2-5 sessions per week. Use the vaginal probe for the treatment of urinary
incontinence in women and the anal probe for fecal incontinence in women
and in men.
URO7-8-9 • Mixed incontinence (medical program)
Program suitable for the treatment of mixed urinary incontinence in women
and faecal incontinence in men (only U7). The stimulation should be as strong
as possible without being painful. In addition, it helps if the patient performs
voluntary contractions during the stimulation phase. To be associated with
appropriate training exercises for muscle strengthening. Applications: 3-5
sessions per week. Use the vaginal probe for the treatment of urinary
incontinence in women and the anal probe for fecal incontinence in women
and in men.
IACER Srl
Description
Mixed incontinence
Mixed incontinence
Total time 20min
Frequency 9s 3Hz + 6s 22Hz
Pulse width 180µs
Total time 20min
Frequency 12s 3Hz + 8s 25Hz
Pulse width 180µs